| System Bus |
Wireless |
| Bus Information |
Bus Standards Compliance |
USB 1.1 USB 2.0 |
System Bus Details |
Wireless: - Stds: IEEE 802.15.4, ISM 2.4GHz frequency band
nonbeacon, point-to-point - Transmit Power Output: 10mW
(10dBm) - Receiver Sensitivity: -100dBm (1% packet error
rate) - RF Channels: Ch. 12-23 (2.410-2.465GHz; software
selectable) - Addressing: 16-bit personal area network addresses per
ch., with 64-bit device addresses, software selectable - Encryption:
128-bit AES, software selectable
Wired Bus: - Stds: USB 2.0, USB 1.1 - USB Type B (4-pin,
square) connector | | Designed to Replace |
UMDAS TC |
| Mounting Configuration |
Desktop / benchtop External module |
| Direct Sensor Input |
Thermocouple |
| On-board Microprocessor |
Yes |
| On-Board Memory for Processor |
1K |
|
| Number of Channels |
8 |
| Differential Channels |
8 |
| A/D Converter Resolution |
24-bit |
| Cold Junction Compensation |
CJC - 2 channels, one left, one right |
| Noise |
CMRR (Common Mode Rejection Ratio) |
fIN = 50 Hz/60 Hz = 100 dB min. |
NMRR (Normal Mode Rejection Ratio) |
fIN = 60 Hz = 90 dB min. | | Input Impedance |
5000MΩ |
| A/D Special Features |
Open thermocouple detect |
|
| Thermocouple Support |
Thermocouple Inputs |
J-type thermocouple K-type thermocouple T-type thermocouple E-type thermocouple R-type thermocouple S-type thermocouple B-type thermocouple N-type thermocouple |
J-type Range Supported |
-346°F....2192°F
(-210°C....1200°C)
| K-type Range Supported |
-346°F....2501.6°F
(-210°C....1372°C)
| T-type Range Supported |
-328°F....1112°F
(-200°C....600°C)
| E-type Range Supported |
-328°F....1832°F
(-200°C....1000°C)
| R-type Range Supported |
-58°F....3214.58°F
(-50°C....1768.1°C)
| S-type Range Supported |
-58°F....3214.58°F
(-50°C....1768.1°C)
| B-type Range Supported |
482°F....3308°F
(250°C....1820°C)
| N-type Range Supported |
-328°F....2372°F
(-200°C....1300°C)
| | A/D Range Selection |
Fixed |
| Input Range Details |
Range is automatically selected by configuring sensor type in software |
| Gains |
1 |
| Isolation, Field-to-Logic |
500VDC |
| Isolation Details |
500VDC min. input isolation between field wiring and USB interface |
|
| Triggering Features |
Triggering is fixed at 2 (24-bit) samples per second on each channel |
|
| Number of I/O Lines |
8 |
| Digital I/O Details |
- Independent configuration for input or output
- Power On Reset: Input mode unless bit is configured for alarm
- All pins pulled up to +5V through 47Kohm resistors (Pulled down to ground available)
- All ground pins (9, 19, 28, 38) are common and earth ground isolated. Connecting this ground to earth ground will make the thermocouples No Longer Isolated.
- Digital Input transfer rate: 50 port reads or bit reads per second typical
- Digital Output transfer rate: 100 port writes or bit writes per second typical
|
| Primary Digital I/O |
Configuration |
8 bidirectional |
Input/Output Selection |
Software | Signal Connector |
Screw terminals | Inputs |
Input Voltage - High (Vih, DC) |
2V....5.5V | Input Voltage - Low (Vil) |
-0.5V....0.8V | Input Current, Low (Iil) |
2.5mA | Outputs |
Output Voltage - High (Voh) |
3.8V....5V | Output Voltage - Low (Vol) |
0mV....0.7V | Output Current Sink, max. |
2.5mA | Output Current Sink Details |
2.5mA @ 0.7V | Output Current Source Details |
-2.5mA @ 3.8V | |
| Software Included Free with Product |
calibration software |
| Software Included - Features |
Driver software |
| Operating System Compatibility |
Windows XP Windows 2000 Windows Me |
| Software Compatibility |
Universal Driver Library DASYLab® LabVIEW® |
|
| Connector #1 |
Quantity |
52 |
Type |
Screw terminals | Location |
On module | | Connector & Cabling Details |
- 52 screw terminals arranged in a block of 10 and of 16 screw terminals on
either side of the module, with a CJC sensor on each side, between each set of blocks
- Wire Gauge Range: 16 to 30 AWG
|
|
| Construction Material |
Industrial plastic |
|
| Power Consumption (typical) |
500mA consumption max (bus powered mode) |
| Power Supply Notes |
2.5-watt USB (type 'A' output) power adaptor with 2-prong US plug adapter |
|
| Other Control Features |
Indicators on the left side of the unit:
- Command LED
- Wireless Power LED
- Transmit LED
- Receive LED
- Received Signal Strength (group of 3 LEDs)
LED Test button is also on the left side of the unit. |
| Display |
LEDs show power and signal status |
|
| Operating Temperature |
32°F....158°F
(0°C....70°C)
|
| Storage Temperature |
-40°F....185°F
(-40°C....85°C)
|
| Relative Humidity, noncondensing |
0%....90% |
|
| Safety/Approvals |
Meets CE standards |
|
| Width |
5.06"
(128.58mm)
|
| Height |
1.51"
(38.33mm)
|
| Depth |
3.48"
(88.37mm)
|
|
| Weight |
0.42lbs
(190.51g
)
|